Experimental Evaluation of a Multi-Pinhole Collimator for a Small Organ by Using a Small-Field-of-View Gamma Camera
Experimental Evaluation of a Multi-Pinhole Collimator for a Small Organ by Using a Small-Field-of-View Gamma Camera
- 주제(키워드) Single-photon emission computed tomography , Experimental data , Multi-pinhole collimator , Collimator modeling , Image reconstruction
- 발행기관 한국물리학회
- 발행년도 2017
- 총서유형 Journal
- UCI G704-000411.2017.70.4.015
- KCI ID ART002201112
- 본문언어 영어
초록/요약
The aim of this study is to design and evaluate a multi-pinhole (MP) collimator for a gamma imaging system that requires a high sensitivity, organ-specific, and small footprint. To ensure these requirements, we designed an eight-hole collimator that can be integrated into a small field-of-view gamma camera for imaging the thyroid or relatively sized organs. Each pinhole was designed to have a cylindrical shape with a 2-mm diameter. Experiments were performed with both a two-sphere phantom and a four-rod phantom. An image reconstruction based on the maximum likelihood expectation maximization with the distance-driven method was used for obtaining a 3-dimensional image. For improving the uniformity of the reconstruction image, we modeled the sensitivity of the cylindrical pinhole by calculating the area of the overlapped circle. The results show that the full width at half maximum values of the two-sphere phantom and the four-rod phantom were 7.56 mm (5-mm-diameter source) and 6.84 mm (5-mm-diameter rod), respectively. The scanning time can be reduced by up to 20 minutes in small-organ applications by using developed MP collimator. Thus, the results indicate that the proposed MP collimator is suitable for a fast scan time, as well as for organ-specific and small-footprint applications.
more